Innovation | Action | Excellence
Flying with Light

CO₂ gas phaseCO₂气相还原

PLR-HGPR High temperature and high pressure photocatalytic reaction system

PLR-HGPR高温高压光催化反应系统

Column:CO₂气相还原Brand:PerfectlightViews:898
PLR-HGPR High temperature and high pressure photocatalytic reaction system
  • Introduction
  • Application
  • Literature
  • Maintenance

Key Features

● Differential photocatalytic reactions at high temperature and high pressure with low gas flow rates;

● Suitable for differential and integral photochemical reactions in gas-solid and liquid-solid phases;

● Independent control of gas flow rates in three channels, utilizing extra-thick quartz windows for photochemical reactions across the full spectrum;

● Equipped with secondary flow control valves and secondary pressure control valves, allowing separate control of three gas inlets.

 

Applications

▲ Particularly Suitable     ● Moderately Suitable   ○ Applicable

▲ Photodegradation of liquid pollutants (e.g., dyes, benzene, and benzene derivatives)         ▲ Photocatalytic CO₂ reduction

▲ Other high-temperature and high-pressure (photo)chemical reactions        

● Photocatalytic water splitting for hydrogen/oxygen production                     ● Photocatalytic total water splitting                                

● Photodegradation of gaseous pollutants (e.g., VOCs, formaldehyde, nitrogen oxides, sulfur oxides, etc.)                                             

○ Photocatalytic quantum efficiency measurement                     ○ Photosynthesis                  ○ PEC (Photoelectrochemical) Reactions

 

Technical Parameters

● Reaction Temperature: Room temperature +10~300 °C (up to 380°C);

● Reaction Pressure: 1~10 MPa;

● Reaction Volume: 50 mL (customizable to 25 or 100 mL);

● Number of Gas Channels: 3;

● Total Gas Flow Rate: 1~60 mL/min;

● Equipped with a safety pressure relief valve (adjustable) of 7 MPa; temperature and pressure indicators for the reaction system;

● Independent high-pressure gas mixing chamber: 1 L, ensuring continuous and long-term gas supply during experiments; equipped with secondary flow control valves and secondary pressure control valves, allowing separate control of three gas inlets;

● End-point soap film flowmeter for gas flow rate monitoring (adjustable in a wide range from 10 to 60 mL/min);

● Overall design with a safety-integrated protective enclosure, ensuring the safety of the experiments;