In order to ensure the repeatability and accuracy of experiments, it is necessary to measure the light irradiance using a light power meter before each photocatalysis experiment. Different models of light power meters should be selected based on the light source intensity and spectral distribution.
The FZ-A type radiometer is a portable, auto-ranging test instrument. It employs SMT surface-mount technology and utilizes high-precision, low-power consumption digital chips. The instrument's exterior features a streamlined design, and the detector undergoes strict calibration for spectral and angular characteristics, ensuring stable performance and high versatility. It can be used to measure visible light irradiance.
● Wavelength Range (λ): 400 nm~1000 nm calibrated using a tungsten lamp, suitable for measuring near AM1.5 G solar spectra (400-1000 nm);
● Irradiance Measurement Range: (0.1-1.999×105) μW/cm2 ;
● Non-linearity, range switching, fatigue characteristics, etc., errors all comply with national first-class illuminance meter standards.
● Accuracy: ±5% (relative to NIM standards);
● Cosine Characteristics Error: f2 ≤ 4%;
● Response Time: 1 s;
● Operating Environment: Temperature (0~40) ℃; Humidity <85% RH;
● Size and Weight: 160 mm (L)×78 mm (W)×43 mm (H), 0.2 kg;
● Power Supply: Typically powered by 1 piece of 6F22 9V laminated battery, can also be powered via a data cable connected to a USB interface or a 5V power adapter, with total power consumption <0.1 VA;
● Probe: FZ-A probe; Diameter: 27 mm; Head Height: 21.5 mm; Photosensitive Hole: 7 mm.